background image

350VL Patient Quick Reference

This guide is meant as a complement to the Operation Manual.
Please refer to the Operation Manual for complete instructions
for use.

Display
Message

Sounds / Lights

Condition

Corrective Action

Sounds: Slow beeping 

Lights: 

Blinking yellow

Not delivering

Press 

 key to start delivery. 

Press 

 key to silence alert for 5 minutes.

Sounds: Slow beeping 

Lights: 

Blinking yellow

Low battery

Press 

 key to start delivery. 

Replace battery within 30 minutes.

Sounds: Continuous tone 

Lights: 

Continuous red

Depleted battery

Remove depleted battery to silence alarm. 
Replace battery. 

Press 

 key to turh pump on. 

Press 

 key to restart.

Sounds: Continuous tone 

Lights: 

Blinking red

Clamp bar is open

Press 

 key to silence alert. 

Close clamp bar. 

Press 

 key to restart.

Sounds: Continuous tone 

Lights: 

Blinking red

IV tubing occlusion

Check tubing system for kinks and correct. 
Pump will resume delivery when line is cleared. 

Press 

 key to silence alert. 

Check tubing system for kinks and correct. 

Press 

 key to restart.

Continued on following page