4

Testing Bracelets

“End of Service” feature

The Model 36000 series has an “End of Service” feature. During routine
testing the Universal Tester will display when the bracelet is reaching
the end of its useful service. This lets you know it is time to replace the
bracelet BEFORE the bracelet becomes inoperable. Bracelets must be
tested daily. IMPORTANT: Failure to replace the bracelet “on time”
could result in injury to or the death of a person in your care.

To test a WanderGuard

®

bracelet with ID capability (36000 series):

Hold the Universal Tester within one foot of the bracelet.

Press the 

BRACELET

button once. The LED above the button will

light briefly.

Press
button

The display will

animate until a

bracelet

is detected, or for

10

seconds.

Summary of Contents for 18029

Page 1: ...arly the Warnings and Cautions before using the WanderGuard Universal Tester Failure to do so could result in injury or death to a person in your care WARNING Operating Instructions WanderGuard Univer...

Page 2: ...copyright notice and is protected by United States and international copyright laws The trademarks service marks and logos the Trademarks used and displayed in this publication are registered and unr...

Page 3: ...ster contains a long life lithium battery that is not user replaceable Do not disassemble heat above 100o C or submerge in water WARNING Dispose of the Universal Tester immediately when low battery is...

Page 4: ...reaching its replace bracelet date The Universal Tester has a non replaceable battery Battery life will vary depending on usage The battery can be expected to last 2 years under typical use The tester...

Page 5: ...DOOR STATUS LED BRACELET STATUS LED The tester has an ambient light sensor built in When the tester is used in low light conditions it will automatically back light the LCD display Using the RECALL bu...

Page 6: ...BEFORE the bracelet becomes inoperable Bracelets must be tested daily IMPORTANT Failure to replace the bracelet on time could result in injury to or the death of a person in your care To test a Wande...

Page 7: ...the bracelet is not detected the LED will be a steady red and display If the Universal Tester does not detect the bracelet Move the tester closer to the bracelet Hold the tester at a different orienta...

Page 8: ...ton will light briefly If the bracelet is detected 000 will be displayed If the bracelet is still operational the LED will flash green 4 times If the bracelet has reached the End of Service phase 000...

Page 9: ...er closer to the bracelet Hold the tester at a different orientation in relation to the bracelet Sweep the tester by the bracelet NOTE Test one bracelet at a time Surrounding bracelets should be at le...

Page 10: ...iversal Tester will test all WanderGuard door monitors except the FirstQTM series Test door monitors at least weekly on each shift with all surrounding power devices turned on to maximize possible ele...

Page 11: ...one located at the doorway The Universal Tester will automatically stop after button is depressed for 30 seconds To test a WanderGuard door alarm or motion detection alarm 1 Open the door s or activat...

Page 12: ...trated below you may have to kneel to keep the tester in the vertical position 2 Press the DOOR button 3 The antenna should chirp and Door Test will be displayed on the System Status Display if instal...

Reviews: