106
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
Technical data
Usable space *
Lara 17 - W: 190 mm/H: 190 mm/D: 310 mm (W: 7.5’’/H: 7.5’’/D:
12.2’’)
Lara 22 - W: 190 mm/H: 190 mm/D: 390 mm (W: 7.5’’/H: 7.5’’/D:
15.36’’)
Bacteriological filter
0.3 µm
*
: usable space with standard rack and trays. With optional racks and trays,
see "Accessories, spare parts, consumables" on page 110.
STEAM GENERATOR
Pressure safety valve
5 bar (72.51 psi)
Safety thermostats
230 °C (446 °F)
DISTILLED OR DEMINERALIZED WATER
Water quality
Fulfilling EN 13060 Ann. C (conductivity: < 15μS/cm, Total
Dissolved Solids: < 10 ppm)
Average water
consumption
0.13 to 0.59 litres/cycle (0.03 to 0.16 gal/cycle)
Tank volume
Clean water 4.8 l (1.27 gal), 2.8 l (0.74 gal) with air gap
Used water 4.8 l (1.27 gal)