Sterilizer tests
42
Lara | Instructions for Use | RIL-1 W&H Med ENG Rev06 | 14/04/2022 | © W&H Sterilization Srl
3
On the homepage, tap
Helix-B&D test
.
4
To set the duration of the Plateau/Sterilization phase and other settings, tap
.
5
Tap and enter your credentials if required: the chamber door locks.
6
Wait until the end of the test and tap
OPEN
: the chamber door unlocks.
7
Enter your credentials if required.
8
Open the chamber door, extract the tray using the tray holder and take the test
pack. The test pack can be wet outside.
CAUTION!
Risk of burns. The test pack is very hot at the end of the cycle. Wear
appropriate PPE (e.g. gloves).
9
Remove the indicator sheet from the center of the test pack and check the change
in colour. See "Interpret the test result" on the next page