20
7. Operation
> Wear protective clothing, saftey glasses, face mask and gloves.
> Refer to the »hygiene and maintenance« instructions in the Instructions for Use of the transmission instruments.
Transmission instruments with ISO coupling
The maintenance program will be completed in about 35 seconds.
> Pull down cover.
Press the green release
button and pull off
the basic adaptor.
Keep the adaptor in
the recess.
With its spray exit nozzle
directed downwards, push
the transmission instrument
onto the maintenance
attachment up to the limit
stop. Close the cover and
the waste disposal tray.
Press the start button once
for about 2 seconds. Open
the cover at the end of the
maintenance cycle.
Press the green release button
of the maintenance
attachment while at the
same time removing the
transmission instrument.
1 x
2 sec.
Maintenance process