22
Operation
Maintenance process
With its spray exit nozzle
directed downwards, push
the turbine with quick release
coupling onto the adaptor until
it engages.
Close the cover and the
waste disposal tray.
Push the appropriate adaptor
onto the maintenance
attachment up to the
limit stop.
Press the start button once
for about 2 seconds. Open
the cover at the end of the
maintenance cycle.
Press the green release
button of the maintenance
attachment, while at the same
time removing the turbine from
the adaptor.
1 x
2 sec.