The available sterilization cycles
42
STERILIZATION CYCLES
Sterilization temperature
134
°
C
134
°
C
121
°
C
134
°
C
Sterilization pressure
3.03 barA
3.03 barA
2.04 barA
3.03 barA
Duration of the holding time phase
4'
18‘ 30”
15‘ 30”
3’ 30”
Duration of the drying phase (ECO-DRY PLUS mode)
5’-16'
5’-16'
6’-24'
1’ 20”
Minimum duration of the drying phase (set by the user)
16'
16'
24'
-
TOTAL CYCLE
DURATION
(1)
(including drying
time)
Lisa 17
Load Empty
Full
Typical
Empty
Full
Typical
Empty
Full
Full
21'
44'
30'
36'
60'
46'
36'
60'
13'
Lisa 22
Load Empty
Full
Typical
Empty
Full
Typical
Empty
Full
Full
21’
49’
31’
36’
64’
47’
36’
63’
15’
LOAD TYPE
Full porous (80% of the usable space).
YES
YES
YES
NO
Small porous items (gauze, cotton, ...)
YES
YES
YES
NO
Hollow A (forceps, scissors, ...)
YES
YES
YES
NO
Bagged, single/double wrapped
YES
YES
YES
NO
Unwrapped items
YES
YES
YES
YES
Full solid (probes, tweezers, burs, ...)
YES
YES
YES
YES
Dental load (turbines and handpieces family, ...)
YES
YES
YES
YES
Hollow B (vacuum tips, ...)
YES
YES
YES
YES
MAX. LOAD
bagged
unbagged
porous
Lisa 17
instruments (bagged): 4.5 kg – porous: 1.5 kg
0.5 kg
(unwrapped)
Lisa 22
instruments (bagged): 6.0 kg – porous: 2.0 kg
0.6 kg
(unwrapped)
MAX. LOAD
containers
Lisa 17 – Lisa 22
Containers: 9.0kg
NOT SUITABLE
(1)
The total cycle time may vary depending on the type of load (solid or porous), the load weight, the duration of the drying phase (if the eco-dry mode is disabled) and other
factors.