50
While a cycle is in progress, you
can abort it manually at any time
by touching the STOP icon.
CONFIRM icon: you confirm that you want to abort the
cycle.
Once a cycle is aborted, the sterilizer starts a reset phase.
Manual stop
A confirmation request appears.
CANCEL icon: you cancel the command. The cycle goes on as
originally programmed.
- the message LOAD NOT STERILE means that the load is not sterile. Do
not use items on patients!
- the message DRYING INTERRUPTED means that the load might be
wet. Wet items are for immediate use only!
Check the WARNING MESSAGE:
By the INFO icon you may display the INFO screen (see RUNNING A STERILIZATION CYCLE)
Do not switch off the sterilizer during the reset phase: it takes some time to reset
the system and reach safe conditions in the sterilizer chamber.
Water could be present in the chamber when opening the door: prevent spilling (e.g., place a towel below the chamber door).
At the end of the reset phase, you may open the door. Reprocess the load, if necessary.