56
7. Cycle history
To access the cycle history: from the CYCLE SELECTION screen (HOME screen), select MENU – TRACEABILITY – CYCLE HISTORY.
This displays the whole list of cycles, identified by number, name, date and time.
Simply touch one cycle in the list, and you have the CYCLE
REPORT displayed.
You may scroll the report list
using the scroll bar and the scroll buttons.
All cycle reports are saved on the USB pen drive.
However, you may copy all cycles on another pen drive by
touching the SAVE ALL icon (a further confirmation will be
requested).
NOTE: only the last 400 cycles will be copied in HTML format, and
the last 50 in SCL format).
See next page for further
options from the CYCLE REPORT
screen.
NOTE: the sterilizer stores in memory the summarized reports of the
last 400 cycles in HTML format, and the analytical reports of the last
50 cycles in SCL format (all parameters recorded every second).
All reports are also saved on the pen drive.
NOTE If for any reason (memory full, pen drive disconnected) some
cycles cannot be saved, no alert is shown.
The unsaved cycles will be copied (provided that they are still stored
in the memory) as soon as a new cycle is started with a working pen
drive connected.