68
ERROR CODE
DESCRIPTION AND ACTIONS
0xx
Load cannot be considered sterile (See END OF STERILIZATION CYCLE).
Repeat the cycle.
If the problem persists, call
service.
Check if the main switch or network circuit breaker is OFF.
Check if the mains cable is properly connected
Switch the sterilizer OFF and ON.
Set date and time, then switch the sterilizer OFF and ON.
Check the dust filter and ensure that the sterilizer fan is not blocked.
10X
See error 13X-16X.
12X
Wait before opening the chamber door. Allow the sterilization chamber to cool down. See error 13X-
16X.
13X-16X
Check water level in the clean water tank. Reset the thermal overload.
Switch the sterilizer OFF and ON.
Clean the door seal and the chamber face side.
Check if the load placed in the sterilization chamber complies with the MAXIMUM WEIGHT LIMITS.
Clean the chamber and the chamber furniture from residuals of detergents, disinfectants and other
chemicals.
Replace the clean water if it is suspected to be contaminated with chemicals.
Ensure all the load is clean rinsed and free from any chemicals before sterilizing.
Start a vacuum test to check the tightness of the pneumatic circuit.
Error table (continues in the next page)