73
The door locking system is electrically activated. In case the door remains locked due to a black-out or an electric fault, an auxiliary
unlocking procedure is available.
1) Extract the dust filter and remove the auxiliary cable loom placed in
the inner side of the handle.
2) Plug two 9V batteries (size PP3 or 1604) into the connectors.
3) Pull the door firmly.
4) Holding the door pulled, plug the plastic connector into the socket
located in the sterilizer, behind the dust filter handle.
5) As soon as the door opens, unplug the battery supply to prevent
system overstroke and consequent damage.
Warning!
The following procedure should be used only if necessary and only when NO RESIDUAL PRESSURE IS IN THE
CHAMBER and the sterilizer has completely cooled down. The sterilizer should be unplugged from the mains
power supply since 3 hours at least.
Any attempt to open the door while the unit is still hot or under pressure, could expose the operator and the
surrounding personnel to serious risk (explosion, jet of hot steam, sudden opening of the door), and could
damage the sterilizer.
Battery-operated door opening tool
1
2
3
4
See video animation on WH.com FAQ videos.