77
CONFORMITY
STERILIZER featuring sterilization cycles conform with the following standards:
93/42/CEE
Medical Device Directive (MDD)
PED 2014/68/EU
Pressure Equipment Directive (PED)
2012/19/EU
Waste Electrical and Electronic Equipment (WEEE)
EN 13060
Small steam sterilizer
IEC 61010-1
Safety requirements for electrical equipment for measurement, control and
laboratory use, general requirements
IEC 61010-2-040
Safety requirements for electrical equipment for measurement, control and
laboratory use; particular requirements for autoclaves using steam for the
treatment of medical materials and for laboratory processes.
EN 61326-1
Electrical equipment for measurement, control and laboratory use: EMC
requirements.
TECHNICAL DATA
Electrical supply:
Nominal voltage:
Max. current:
200 - 240 VAC – 50/60 Hz, single - phase
10 or 12 A (refer to product label)
Sterilizer:
Working temperature:
Working relative humidity:
Storage temperature/ rel. humidity:
Max altitude:
Min. atmospheric pressure:
Overall dimensions:
Min. space required:
Size of the door movement:
Weight empty:
Max. weight (fully loaded):
Weigh per support area
Max. heat output:
from +5
°
C to +40
°
C
Max. RH 80% up to 31
°
C, linearly decreasing to 50% at 40
°
C
-20
°
C to +60
°
C/0-90% (with empty tanks)
3000m asl
0.6 barA
W: 465 mm/H: 449 mm/D: 599 mm
W: 485 mm/H: 499 mm/D: 619 mm
W: 410 mm/H: 400 mm/D: 360 mm
50.2 kg (LISA 17), 51.7 kg (LISA 22)
65.3 kg (LISA 17), 66.8 kg (LISA 22)
38.7 kN/m
2
(LISA 17), 39.8 kN/m
2
(LISA 22)
3000 KJoule/hour
Sterilizer chamber:
Pressure safety valve:
Safety thermostats
2.6 bar
180
°
C
Steam generator
Pressure safety valve:
Safety thermostats:
4,0 bar
230
°
C
Sterilizer chamber:
Total volume:
Usable space (for all cycles)
Bacteriological filter:
17 l/Ø 250 mm x D 362 mm (LISA 17)
22 l/Ø 250 mm x D 440 mm (LISA 22)
11.5 l/W 195 mm x H 195 mm x D 297 mm (LISA 17)
15 l/W 195 mm x H 195 mm x D 390 mm (LISA 22)
0.3 µm
Distilled or demineralized water:
Water quality:
Average water consumption:
Tank volume:
Fulfilling EN 13060 Ann. C (conductivity< 15μS/cm)
0.32 to 0.65 litres/cycle
Claean water 4 l – Used water 3.5 l
External water supply:
Pressure:
Flow:
Temperature:
To be compliant with IEC61770
min. 2 bar – max. 8.6 bar
min. 0.25 – max 0.5 l/min
Max. 35
°
C
Communication with other devices:
1 Serial port - 3 USB port - 1 Ethernet port
Other
Fully micro-processor controlled, process evaluation system
according to EN13060.
Programmable sleep-mode.
ANNEX 1. Technical data
LISA sterilizers can be validated in accordance to EN17665-1.
RECOMMENDATIONS FOR VALIDATION
Coldest point
Hottest point
LISA 17
LISA 22