4. Installation, programming and setup
9
Required minimum clearances
Back side:
25mm (defined by the protrusion in the back)
Right and left sides:
10 mm
Upper side:
As required for filling the water tank, 50 mm minimum
Placement
Place the sterilizer on a flat and level surface, far from sources of heat and from flammable materials.
Do not place the sterilizer so that it is difficult to open the service door and operate on the controls behind it.
Do not place the sterilizer so that it is difficult to disconnect the power supply plug.
Place the sterilizer in a well ventilated room.
The sterilizer must not be operated in presence of explosive atmospheres.
Electrical connection
The electrical power supply to the sterilizer must fulfil all applicable standards in the country of use, and must comply
with the data label on the back of the sterilizer.
Connect the cord set to the socket provided in the back of the sterilizer.
Use only the cord set provided by the manufacturer.
The power plug must be connected to a wall socket complying with the following specifications (refer to product label).
Specification
Version labelled 10A
Version labelled 12A
Voltage and frequency
Single phase, 200 - 240 V, 50/60 Hz
Max. current
10 A
12 A
Overvoltage category
II
Automatic circuit breaker (min.)
10 A
12 A
Residual-current device
30 mA or less, if required by local regulations
Ground connection
Required
All protection and safety devices must be certified according to applicable standards.