Electrical connection
Electrical connection
The electrical power supply to the sterilizer must fulfil all applicable standards in the country of use,
and must comply with the data label on the back of the sterilizer.
Use only the cord set provided by the manufacturer.
Connect the cord set to the socket provided in the back of the sterilizer.
Do not use cable extensions nor multiple sockets/adapters
Connect the mains cable to a wall socket with the following characteristics:
•
Single - phase 200 - 240 V, 50/60 Hz - 10A, on a dedicated circuit;
•
Overvoltage category = II;
•
Protected by a suitable circuit breaker and a residual-current device. All protection devices must be
certified according to applicable standards;
•
Maximum current absorbed by the sterilizer: 10A;
•
A grounded connection is essential.