28
The available sterilization cycles
NOTES (1) Cycle names could be different depending on country requirements.
(2) Values could be different depending on country requirements.
(3) The total cycle time may vary depending on the type of load, the load weight, and other factors.
(4) The evaporation time can be increased by the SETUP menu (see “Customization of cycle parameters“) if required.
(5) Moisture acceptance criteria according to ANSI/AAMI ST55.
(6) Bagged /wrapped items shall be put in the upper tray only.
(7) With aluminium trays.
CYCLE PROGRAM TABLE
CYCLE
PROGRAM
NAME
(1)
LOAD
Sterilization phase
Total cycle time
Minutes
(3)
Evaporation
time
(4)
minutes
Type
Max.
Weight kg
Temperature
°
C
Time
(2)
minutes
Warm
Start
Cold
Start
Dental load as defined in ISO 17665-3
(7)
.
W&H product families including
handpieces and turbines. UNWRAPPED.
2
134
3.5
18'11" 21'30"
2
(5)
Dental load as defined in ISO 17665-3
(7)
.
W&H product families including
handpieces and turbines.
WRAPPED
(6)
.
2
134
3.5
58'
62'
40
(5)