Resetting the thermal overload
A safety thermostat is fitted on the sterilizer to prevent overheating
of the electric heater.
If the safety thermostat opens because of too high temperatures,
the alarm E240 or a timeout alarm is generated.
If this happens, proceed as follows:
- Switch the sterilizer OFF and remove the mains cable.
- Wait for the sterilizer to cool down.
- Remove the dust filter.
- Slide your hand underneath the sterilizer where the dust filter
was located and push on the reset button of the thermostat switch
(see pictures to the left).
-A click sound will indicate that the thermostat switch has been
reset.
- Insert the dust filter back into its original position.
-Connect the mains cable and switch the sterilizer ON.
- Wait for the sterilizer to finish the alarm reset phase and follow
the instructions on the display.
If the thermostat opens repeatedly, call technical service.