USE OF OSMO
Once the system has been correctly installed and connected, the demineralizer is able to immediately supply purified and demineralized
water.
If the demineralized water is used to run a sterilizer, you may choose between manually filling the sterilizer tank with the water dispenser, or
having it filled automatically (if the sterilizer is suitable for connection to a purified water system).
The quantity of water produced by the filter cartridges depends on the quality and hardness of the city water supply in the installation area.
PRECAUTIONS TO AVOID WATER CONTAMINATION
7. Use and operation
21
In accordance with subsection 4.5 of EN1717, in order to avoid water quality degradation and bacterial growth, the entire
system must be drained before leaving it unused for more than 10 days.
Re-using the system after a long period of non-use.
To restart the equipment when it hasn’t been used for a long time or after the equipment has been emptied, the system must
be flushed for at least 5 minutes (see section 6 “Priming”, points 1, 2 and 3).
It is also recommended not to leave the storage tank filled with demineralised water for a long period of time.
BY-PASS ADJUSTMENT
The pump is equipped with a by-pass system to adjust the fixed residue of the produced water.
Don’t change the by-pass system regulation.