15
Safety notes
Control unit PB-510
> Disconnect the medical device from the water supply connection after each use (the medical device does not have an
automatic aquastop).
> The medical device is approved for use only with supply units with category 5 backflow prevention devices as defined
in EN 1717.
> Do not connect the medical device to the hot water supply (>30°C/86°F).
Control unit PB-520, PB-530
> Never fill or top up the coolant tank with liquids hotter than 30°C/86°F.
> Replace a faulty or leaky pump cartridge immediately.
Control unit PB-510, PB-520, PB-530
Risks due to electromagnetic fields
This medical device complies with the reference values defined in EN 50527-2-1/2016 for active implantable medical
devices (AIMD) and with cardiac pacemakers.
The control unit is designed for use with the W&H handpiece PB-5 L/L S/L Q so only this is to be used with the control unit.
The use of other handpieces could lead to a malfunction or destruction of the electronics.
Control unit