General Maintenance 

 

 

It is strongly recommended that the following checks be carried out on a weekly basis by facility 

personnel to ensure trouble free operation of the MR4500 folding portering chair. 

 

 

 

Check that the portering chair runs straight (no excessive drag or pull to one side). 

 

 

Ensure that there is no excessive side movement of rear wheels when the portering chair is 

lifted and wheels spun. 

 

 

Inspect seat and back upholstery for sagging, cracking and any other obvious signs of  

damage or contamination.  Report any damage. 

 

 

Inspect hand grips for wear, looseness and / or deterioration. 

 

 

Inspect for any broken or loose parts. 

 

 

Inspect for damage to tyre and wheels generally.  Remove any debris that may have        

become caught in the brake and around the axle, this is a common occurrence, and can 

lead to a shortened life expectancy of the castors if left unattended. 

 

 

Check for correct operation of brakes. 

 

 

Check castors for free movement and rotation, remove any debris caught in castor. 

 

 

Check armrest for correct operation and damage to padded section. 

 

 

Check operation of footrest including possible damage to heel straps. 

 

 

Do not attempt to rectify any faults on the chair without contacting Wardray Premise. 

 

There is no Service Manual available for this product. 

 

Please register your product with us, via the Warranty Registration Form (see page 15), for ease 

of contact in case of urgent upgrades or recalls. 

 

If the unit is damaged please do not attempt to repair as this could invalidate your 

warranty and could compromise its MR integrity and therefore safety, please call or 

email Wardray Premise Limited: 

 

T: +44 (0) 20 8398 9911 

F: +44 (0) 20 8398 8032 

E-MAIL: sales@wardray-premise.com 

 

Please include the product code and the product’s serial number when 

contacting Wardray Premise. 

 

Summary of Contents for MR4500

Page 1: ...GEN S 17 175 July 2015 MR4500 MR Conditional Folding Portering Chair OPERATORS MANUAL GEN S 19 175e DECEMBER 2019...

Page 2: ...ormation regarding operation service and maintenance If you have any problems in the meantime or would like any advice about this or any other MR products from the Wardray Premise range please contact...

Page 3: ...ned conditions At a minimum address the conditions of the static magnetic field the switched gradient magnetic field and the radiofrequency fields Additional conditions including specific configuratio...

Page 4: ...ded strength It is advisable that this product is inspected and serviced every 12 months by a Wardray Premise trained technician A Service contract package is available and details are available upon...

Page 5: ...0 Portering chair parts reference Footrests Castor Adjustable armrest Front Brake Rear Brake Rear Wheel Plungers Seat Backrest In folded position Push handles Threaded Lobe Knob Serial No located on u...

Page 6: ...to a position below the seat level This can easily be achieved by pulling the plunger to disengage then lower the armrest To raise the armrest simply pull the armrest upwards It will click as the plu...

Page 7: ...e footrest away from the chair in a horizontal motion removing the footrest and the supporting arms To refit the footrest simply align the two supporting arms and push the footrest back into place Onc...

Page 8: ...and push down firmly the brake will lock in place and stop the castor from moving To release the brake place your foot on the rear part of the brake lever Fig 2 and press down firmly this will releas...

Page 9: ...the seat The chair will assume the folded position Fig 2 There is no locking mechanism the weight of the chair keeps it in the folded position To open the chair from its folded position stand behind t...

Page 10: ...erly trained staff at all times The portering chair must only be used for the purpose for which it is designed Always refer to and comply with your organisation s patient movement policy Ensure that p...

Page 11: ...harm may occur attempt to move up or down inclines with water ice or oil film stand on footrest when getting in out of the portering chair make sure they are folded out of the way transfer patients i...

Page 12: ...is product Recommendations Always refer to and comply with your organisations patient movement policy We recommend that all staff using the chair receive appropriate manual handling training and that...

Page 13: ...and can lead to a shortened life expectancy of the castors if left unattended Check for correct operation of brakes Check castors for free movement and rotation remove any debris caught in castor Chec...

Page 14: ...If required the product can be washed down with hand hot sodium hypochlorite solution up to 1 or 10000 parts per million chlorine combined with a detergent Alternatively follow your local guidelines o...

Page 15: ...15...

Page 16: ......

Reviews: