Figure 8.
Low Temperature Light
Temperature Control
WARNING
The patient must be closely monitored for rewarming. Vasodilation and
possible hypotension can occur. Use good judgment when selecting a
temperature. If unsure of proper setting, consult with the attending
physician.
WARNING
Using the warming system on transdermal medication patches may increase
the rate of drug delivery, potentially causing harm to the patient.
The control panel features a warning light indicator and four manually-switched
temperature settings. The warning light indicates the control system has
identified an alarm condition and an action by the operator is required. Each
temperature setting represents the average temperature of air entering the
hose.
•
Low
32°C (89.6°F)
•
Medium
38°C (100.4°F)
•
High
43°C (109.4°F)
•
Boost
45°C (113°F) for 45 minutes
For patients with severe hypothermia requiring rapid warming, select the Boost
temperature setting. To help prevent hypothermia or for treatment of mild
hypothermia, select High, Medium or Low temperature settings. The selected
temperature light illuminates.