Routine Maintenance
Overview
This chapter describes the steps required to maintain, service, and properly
clean the WarmTouch™ Model WT-5900 patient warming system.
Cleaning the Warming System
Caution
Do not spray, pour, or spill any liquid on the warming system, its accessories,
connectors, switches, or openings in the case.
For surface cleaning and disinfection of the warming system, follow your
institution's procedures or the recommended actions below.
•
Surface cleaning
— Use a soft cloth dampened with either a commercial,
nonabrasive cleaner or a solution of 70% alcohol in water, lightly wiping
the surfaces of the warming system.
•
Disinfection
— Use a soft cloth saturated with a solution of 10% chlorine
bleach in tap water, lightly wiping the surfaces of the warming system.
Routine Maintenance
Caution
The institution should follow local governing ordinances and recycling instructions
regarding disposal or recycling of filter and device components or end of life of the
product.
Caution
The HEPA filter
must
be changed every 2,000 hours of operation. Refer to the
Routine Maintenance
section in the
Service Manual
for replacement procedures
requiring a qualified technician.
For routine maintenance procedures requiring a qualified technician, including
testing and verifying operation of the independent over-temperature safety
system and the subsequent over-temperature alarm, refer to the
Routine
Maintenance
section of the
Service Manual
.
Summary of Contents for WT-5900
Page 1: ...Model WT 5900 Patient Warming System Operator sManual ...
Page 24: ...Page Left Intentionally Blank 24 Operator s Manual Using the Warming System ...
Page 26: ...Page Left Intentionally Blank 26 Operator s Manual Routine Maintenance ...
Page 32: ...Page Left Intentionally Blank 32 Operator s Manual Specifications ...
Page 34: ...Page Left Intentionally Blank 34 Operator s Manual ...
Page 35: ......