Safety Information
Operator’s Manual
7
WARNING
WarmTouch
™
CareDrape
™
and CareQuilt
™
blankets are for single patient
use only.
WARNING
The warming system should not be operated in the presence of
electromagnetic fields that are greater than 3 volts/meter. This could cause
shutdown of the warming system by the fail-safe function within the
equipment.
WARNING
The warming system is not suitable for use during magnetic resonance
imaging (MRI) scanning. The warming system may affect the MRI image.
WARNING
Continuously monitor the patient's temperature. Reduce the air temperature
or discontinue therapy when normothermia is reached.
WARNING
The patient must be closely monitored for rewarming. Vasodilation and
possible hypotension can occur. Use good judgment when selecting a
temperature. If unsure of proper setting, consult with the attending
physician.
WARNING
The use of accessories and power cables other than those specified may
result in increased emission and/or decreased immunity of the warming
system.
WARNING
Thermal injury may occur if the warming system hose comes into contact
with the patient.
WARNING
Using the warming system on transdermal medication patches may increase
the rate of drug delivery, potentially causing harm to the patient.
Cautions
Caution
Federal (U.S.A.) law restricts the use of the warming system to sale by or on the order
of a physician.
Summary of Contents for WT-5900
Page 1: ...Model WT 5900 Patient Warming System Operator sManual ...
Page 24: ...Page Left Intentionally Blank 24 Operator s Manual Using the Warming System ...
Page 26: ...Page Left Intentionally Blank 26 Operator s Manual Routine Maintenance ...
Page 32: ...Page Left Intentionally Blank 32 Operator s Manual Specifications ...
Page 34: ...Page Left Intentionally Blank 34 Operator s Manual ...
Page 35: ......