Operation
19
Figure 11- Pressure Setup
Select
Accept
to return to the
Configuration
screen (Figure 9). Before the
method is saved choose a
Canister End State
. The canister end state allows the
user to select if the canisters will be under vacuum or under pressure at the end
of the cycle. If the canisters will need to be tested for certification after cleaning,
select
Pressure
. The pressure of the canister will be left at the
Charge Pressure
of the last cycle. If the canisters will need to be readied for sampling then select
Vacuum.
The canisters will be left at the
Vacuum Pressure
of the last cycle.
This pressure is set in the
Pressure Setup
screen (Figure 11).
After the method has been configured the user can save and name the method
using the keyboard provided in the
Save Method
screen (Figure 12) by choosing
Save
from the
Configuration
(Figure 9)
screen. Type the name of the new
method and choose
Save.
After saving the method the software returns to the
Configuration
screen. From the
Configuration
screen choose
Accept.
This will
return the user to the
Start Menu
(Figure 8)
.