background image

4

INTRODUCTION

The SST Dock is an automated test station that allows Bump and Calibration testing and is com-

patible with the full SST Range in one module. It has the capabilities to be connected via Wi-Fi, BLE 

or Cellular and allows for desk, wall and hard case mounted dependent on your application. With 

two inlets it allows for calibration of the standard four gases plus another gas type. With Built in 

NFC it is easier than ever to assign gas cylinders to the correct gas in ports. 

1

SST DOCK OVERVIEW

1. Touch screen

2. Touch screen ON / OFF

3. Status LED

4. Instruction label

5. Bottle NFC reader

6. Lid lock

7. Power supply

8. Mounting bracker table / wall mount

9. Exhaust port

10. Port 2

11. Port 2 pressure sensor inlet 

12. Port 1

13. Port 1 pressure sensor inlet

14. Fresh air inlet filter

15. Fresh air inlet pressure sensor

16. ON / OFF key

2

3

4

5

6

7

8

9

10

12

11

13

16

14

15

Figure 1.

Summary of Contents for SST Dock

Page 1: ...SST Dock Quick start guide SST Dock QSG V 1 0 EN 28 04 23...

Page 2: ...s office or trained service technician Ensure calibration is the correct quality and that cylin ders are with in date and contain gas Failure to do so may result in an incorrect calibration Warranty d...

Page 3: ...y connect up to 10 meter of tubing on fresh air inlet and outlet Ventilation the outlet should be connected to a ventilated area Adjust the fresh air purge time accordingly Ensure tubing and cables ar...

Page 4: ...on of the standard four gases plus another gas type With Built in NFC it is easier than ever to assign gas cylinders to the correct gas in ports 1 SST DOCK OVERVIEW 1 Touch screen 2 Touch screen ON OF...

Page 5: ...area and to select the proper mounting hardware before mounting the SST Dock Mounting the SST Dock must be performed by a qualified installation contractor SWITCH ON THE SST DOCK If the SST Dock is fu...

Page 6: ...se shall be activated or the SST Dock will ignore the signal Enter the password 2 Add calibration gas bottle in the Gas bottle list either manual or utilize Watch Tab enabled bottles Its important to...

Page 7: ...target gas 2 Status LED Green normal operation and on charge yellow running on battery red fault 3 Server setup linking to the WatchGas compliance software the server ID must be set up This link will...

Page 8: ...on the SST Dock by using the two finger guides and the red latch Step 3 Insert device by locating the alligator clip into the guide in the SST Dock Step 4 Close Lid until you hear the click Step 5 Se...

Page 9: ...Data Storage More than 1 year of operation Connection capability Wi Fi Cellular BLE Connected into the Cloud and WatchGas App Gas Inlets Fresh air inlet 2 Gas inlets and one purge Testing Function Bu...

Page 10: ...ck in liquids Completely turn off unit and disconnect power supply before cleaning PARTS AND ACCESSORIES Replaceable parts and accessories are sold separately Please contact WatchGas or an authorized...

Page 11: ...1 Redo the test 2 Manually inspect the detector for LED vibration and audio alarms 3 Test a known fully functional detector in the station to verify station function Station fail 1 Clear fault and ret...

Reviews: