Section 9 – Custom Settings Menu
41
7. The pH measurement of the
Chemflo
is now calibrated. The
Chemflo
will now prompt you to
re-fit the probe into the Tee fitting…
Re-Fit Probe
into Pipe Enter
8.
Re-fit the probe into the Tee fitting, as per section 2.2.
Return the pool pump to its normal Pump Timer setting, as per section 3.4.
Press
to return the
Chemflo
to Automatic dosing when finished calibrating.
9.7.2 pH Calibration Notes
1. The
Chemflo
has an allowable Asymmetry range of –1.00 to +1.00 pH. If calibration fails due
to the Asymmetry being outside these limits, please consult the Troubleshooting guide (section
7.4) for possible remedies.
2. All calibration information is retained in memory when the
Chemflo
is switched off, even when
the power supply is removed.
9.7.3 pH Calibration Messages
1. If a 7.00 pH has been successfully performed, the
Chemflo
will display the following message
and the asymmetry of the probe. Note that the slope value from the last calibration is also
shown. For example…
pH7 Cal. OK
Asym=0.10pH
2. If a 7.00 calibration has failed, the
Chemflo
will display the following message and the failed
asymmetry value of the probe. For example…
pH7 Cal. Failed
Asym=1.10pH