28
Preparing the Separations Module for Operation
• You change the seal pack
• You rebuild the seal pack
Before you adjust the seal pack, purge the sample management system (
Purging the Sample Management System
) to ensure that there is no air in the syringe. Air
in the system may cause Alarm Seal Geometry or Alarm Missing Restrictor dialog boxes
to erroneously appear.
To adjust the seal pack:
1. Press the
Diag
screen key in the Main screen to display the Diagnostics screen.
2. Press the
Adjust Seals
screen key. The adjustment procedure begins. When the
test is complete, the results are shown in the dialog box that appears.
2.4.4 Loading Carousels
Loading carousels involves:
• Loading the vials into the carousel
• Loading the carousel into the sample compartment
Loading Vials
Use only 12
×
32 mm, 2-mL vials with snap-on, crimp-on, or screw-on caps.
Insert the prepared vials into the appropriate positions in the carousel(s). Be sure to place
vials in the numbered slots that correspond to the sample set (if any) you are using. If you
are not using a sample set, enter the sample vial positions in the Inject Samples dialog
box. If you are using the bar code option, make sure you have correctly attached labels to
the vials.
Loading the Carousel
lists the carousels and their color codes.
Table 2-2 Carousel Codes and Colors
Code
Color
Vial Numbers
A
Blue
1 to 24
B
Yellow
25 to 48
C
Red
49 to 72