Figure 3–80: Inlet manifold outlet-tubing fitting
1
2
3
Inlet manifold outlet-tubing fitting
Inlet manifold
i
2
Valve
6.
Reattach the outlet-tubing fitting to the inlet manifold, and tighten it with your fingers to the
extent possible.
Requirement:
Ensure that the outlet-tubing is fully engaged by raising the inlet manifold
while attaching the outlet-tubing fitting.
Figure 3–81: Reattaching outlet-tubing fitting
1
2
Outlet-tubing fitting
Inlet manifold
7.
Tighten the outlet-tubing fitting, on the side of the
i
2
Valve
with your fingers, and then use
the 1/4-inch open-end wrench to tighten the fitting an additional 1/6-turn, for existing
fittings, or 3/4-turn, for new fittings.
August 8, 2016, 715005050 Rev. C