Figure 2–13: Leak sensor reservior
Leak-sensor reservoir
8.
With a cotton swab, absorb any remaining liquid from the corners of the leak-sensor
reservoir and its surrounding area.
9.
Align the leak sensor’s T-bar with the slot in the side of the leak-sensor reservoir, and slide
the leak sensor into place.
Figure 2–14: Aligning T-bar with slot
T-bar
Slot in leak-sensor reservoir
Leak sensor port
10. If you detached the connector from the front of the device, reattach it.
August 8, 2016, 715005050 Rev. C