To replace the low-pressure inlet filters on the GPV outlet:
Notice:
To avoid damaging electrical components and circuitry, do not
disconnect an electrical assembly while electrical power is applied to a module.
To completely interrupt power, set the on/off switch to the "off" position, and then
disconnect the power cord from the ac source. Wait 10 seconds thereafter before
disconnecting an assembly.
1.
Power-off the solvent manager.
2.
Open the solvent manager’s door.
3.
Disconnect the four inlet manifold tubes from the GPV outlets.
Figure 3–14: GPV outlets
1
GPV outlets
4.
Remove the low-pressure inlet filters from the inlet manifold tubing.
Tip:
If the filters are not on the ends of the inlet manifold tubing, use the tubing to retrieve
them from the GPV outlets.
Figure 3–15: Low-pressure inlet filters on inlet manifold tubing
1
2
Inlet manifold
Low-pressure inlet filters