1-9
Heating and trapping module
The heating and trapping module is attached to the sample manager and
serves as its top cover. The heating and trapping module’s column tray can
accommodate a nano tee and any analytical Waters column up to 300 microns
internal diameter and 250 mm length. The column compartment heats to
temperatures from 5 °C (9 °F) above ambient to 65 °C (149 °F).
To reduce dispersion associated with dead volume and minimize the length of
tubing between system modules, the column tray swings outward to any
position between 0° and 180°. In the 0° “home” position, the column tray is
directly above the sample manager and can be connected to an optional optical
detector. In the 180° “away” position, the analytical column can be plumbed
into a mass spectrometer (located on the system’s right).
Loop: 20
µ
L
Needle: 15
µ
L
Partial loop
5
5
19
Partial loop with
needle overfill
5
9
15
Full loop
Auto
10
40
20
1
×
Overfill
10
20
20
2
×
Overfill
10
40
20
3
×
Overfill
10
60
20
a. Maximum injection volume is approximately 0.1
µ
L
less than actual sample loop volume. Sample
loop volumes are nominal value only; volumes vary slightly from loop to loop.
Sample consumption guidelines (Continued)
System
Configuration
Injection Mode
Requested
Sample Size
(
µ
L)
Sample
Consumed
(
µ
L)
Maximum
Injection
Volume (
µ
L)
a