5-44
Maintenance Procedures
18. Using the combined 2.5-mm Allen wrench and cone extraction tool,
remove the captive screws securing the 2 PEEK extraction cone retainer
clips.
19. Remove the extraction cone from the ion block.
Caution:
• Take great care not to damage the extraction cone aperture
when removing the extraction cone from the ion block.
• The extraction cone is fragile. Never place it on its tip; always
place it on its flanged base.
Securing screw
Extraction cone
Retainer clip
Extraction cone aperture