Replacing the ion block source heater
5-87
Replacing the ion block source heater
Replace the ion block source heater if it fails to heat the ion block.
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
Combined 2.5-mm Allen wrench and cone extraction tool
•
Ion block source heater
To replace the ion block source heater
1. Remove the ion block assembly from the instrument (see
).
2. Ensure that the isolation valve is closed.
Warning:
The ion block assembly can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Isolation valve handle
in closed position