Configuring for APCI mode
3-17
•
At the diverter valve, use a long "finger tight" PEEK fitting.
•
At the probe, use a PEEK nut and ferrule, finger tightened, to
connect to the PEEK union.
Tubing connection between the diverter valve and the APCI probe
(The other tubing connections are omitted for clarity.)
6. Slide closed the instrument’s source interface door.
Caution:
Ensure that the tubing does not become trapped when
closing the source interface door.
APCI probe
Diverter valve
Tubing connection
Probe adjuster
assembly
Long "finger tight" PEEK fitting
PEEK nut and ferrule