4-8
Configuring the NanoLockSpray source
Optional glass capillary sprayer
The glass capillary sprayer is designed for use with metal-coated borosilicate
glass capillaries, which allow extremely low flow rates (less than 100 nL/min).
The capillaries are used for one sample only and must then be discarded.
To use the glass capillary sprayer, you must complete the following
procedures:
•
Install the glass capillary sprayer.
•
Fit and load the glass capillary.
•
Optimize the sprayer.
Installing the glass capillary sprayer
Required materials
Chemical-resistant, powder-free gloves
To install the glass capillary sprayer
1. Prepare the instrument for work performed on its source (see
).
2. In the Tune window, click Source Standby
, and confirm that the
adjacent status display is yellow.
3. Retract the sprayer platform adjuster assembly from the source (see
.
4. Remove the sprayer’s safety cover.
f
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, prepare the instrument for work
performed on its source before commencing this procedure.
Warning:
The probe and source can be hot. To avoid burn injuries,
take great care while working with these components.
Summary of Contents for SYNAPT G2
Page 18: ...xviii Table of Contents...
Page 46: ...2 8 Starting Up and Shutting Down the Mass Spectrometer...
Page 66: ...3 20 Configuring the LockSpray Source...
Page 228: ...B 24 External Connections 7 Click Next Result The software installs 8 Click Finish...
Page 232: ...C 4 Materials of construction and compliant solvents...