Emptying the nitrogen exhaust trap bottle
5-19
Emptying the nitrogen exhaust trap bottle
Inspect the nitrogen exhaust trap bottle in the instrument exhaust line daily,
and empty it before it is more than approximately 10% full.
Nitrogen exhaust trap bottle
Required materials
Chemical-resistant, powder-free gloves
TP03164
From instrument
exhaust connection
To laboratory
exhaust port
Cap
Nitrogen exhaust
trap bottle
From instrument
pilot valve port