background image

December 11, 2015, WAT051512 Rev. B

Page 2

Required materials

From the Waters Gradient Mixer kit, part number WAT051518:

Installing the gradient mixer

To install the gradient mixer: 

1. Position the gradient mixer on the bench top to allow easy visual leak detection. 
2. Connect the gradient mixer as shown in Figure 1:1 

“Gradient mixer 

connections”

Note: 

If you are using 3 pumps, add another tee fitting (not included in the kit) 

as shown in the inset. 

Figure 1–1:

Gradient mixer connections

3. Be sure to use an in-line filter (part number WAT035190 included in the kit) 

since particles in the solvent can clog the 0.005-inch passages inside the 

gradient mixer.

Note:

 If you are operating a ternary pumping system, a second tee fitting (part 

number WAT075215) is required which is not supplied in the mixer kit. 

Table 1–1:

Gradient mixer kit materials description

Description

Waters part number

Quantity

Gradient mixer

WAT051519

1

In-line filter (2 micron)

WAT035190

1

Replacement filter element WAT088084

1

Jumper tube

WAT038148

1

Compression screw

WAT025313

2

Ferrule

WAT022330

2

Tee fitting

WAT075215

1