Watlow D4T Data Logger
•
50
•
Chapter 3 Using the D4T Front Panel
Note:
Cycling the power to the data logger after step two will also initiate an immediate up-
date.
After executing step 6 above, the three screen shots below will appear on screen as the flash-
ing activity proceeds.
Copy files:
File Update:
File Update Complete:
When this last screen appears, it will remain on screen for approximately 15 seconds.
Note:
The data logger will be restored to reflect the same state it was in when the flashing
activity was started. For example, if the data logger was in auto mode it will be returned
to auto mode.
Note:
After the data logger has been updated, the USB Flash Drive needs to be removed from
the data logger. If it remains, when the data logger power is cycled, the firmware flash
will occur again.
Step 1 of 3 : Moving file froom USB flash drive to F4T...
Please do not disconnect USB flash drive or system
power until update completes.
Step 2 of 3 : Updating...
Please do not disconnect USB flash drive or system
power until update completes.
Step 3 of 3 : Update complete.
Please remove USB flash device from F4T.