Chapter 4 Calibration
Using Composer™ Software to Calibrate Analog Inputs
To obtain access to the calibration screens:
1. Start Composer software if not already running.
2. From the Dashboard click on
Connect to a System
and select the appropriate device.
3. From the System Overview screen click on the
Device
menu button where a drop down
list will appear.
4. Click on
Calibrate
.
5. Select the appropriate Module and Input and then click on
Perform Field Calibration
.
6. Follow instructions on the screen.
Using the User Interface to Calibrate Analog Inputs
To calibrate from the F4T front panel:
1. Push the
Menu
,
Service
and
Calibration
buttons, in that order.
2. Select the desired module and input and then push
Perform Field Calibration
.
3. Follow instructions on the screen.
Note:
At any point in time (using Composer software or the UI) the selected module and input
can be brought back to the factory calibration settings by selecting "Restore Factory Cali-
bration".