Watson-Marlow 520DuN IP66/NEMA 4X User Manual
85
The spring-loaded tube clamps must grip the tubing tightly enough to stop it
moving in and out of the pumphead but must not over-squeeze the tube and
throttle fluid flow. The tubing clamps are fitted with yellow sliders which can be
clicked into two positions while the clamps are held open: the outer position
will allow the clamps to grip the tube tightly; the inner will grip the tube loose-
ly. Adjust the sliders to prevent tube movement during a few trial rotations of
the rotor.
Close the guard, pushing it fully home until the latch engages.
Connect suitable pipework to the pumphead tubing using appropriate connec-
tors.
Remember, when using Marprene or Bioprene tubing, re-tension the tubing
after 30 minutes of running, as it may grow in length as it beds in. Re-tension
so that 225mm (8
7
⁄
8
in) of tubing sits between the inside faces of the spring-
loaded parts of the tube clamps.
29.3 520RE: fitting the drain port
The drain port is an optional extra, supplied with the pumphead. It is strongly rec-
ommended that users fit it before the pump is operated. It may be done with the
rotor in position or removed.
Remove the drain plug from the bottom of the pumphead. The drain plug is
flexible. It may be removed using finger pressure from within the pumphead,
or by accessing its flange from outside with a fingernail.
Drop the port into position from inside the pumphead.
Fit the supplied port retaining nut (
3
⁄
8
in BSP) and finger-tighten.
Fit drainage pipework as required (not supplied).