CARDIO WAVE EXCITE Class & Trend: Service & maintenance manual- rev.2.0
Pagina 9.32
9.5.1.4.
Table of configuration parameters
To correctly display the parameter values, you need to load them from the low kit, using
the “Read from low kit” function.
To change the setting, press the
CHANGE
key and insert the value using the numeric keyboard.
Press
ENTER
key to confirm or
BACK
key to back forward.
After any changes to the parameter values, you need to load them in the low kit using
the “Write to low kit” function.
Parameter
Description
Default values
Par 01
Transmission ratio
4248
Par 02
Speed ratio (integer part)
400
Par 03
Minimum braking current (standby pos.) [mA]
500
Par 04
Maximum continuous power [W]
200
Par 05
Machine torque correction coefficient (constant torque)
85
Par 06
User friction torque
10
Par 07
Maximum coil temperature [°C]
125
Par 08
Minimum revolutions [RPM]
35
Par 09
RPM integration scale
400
Par 10
Filter threshold [RPM]
400
Par 11
Speed ratio (decimal part)
0
Par 12
User friction torque delta
30
Par 13
NOT USED
48
Press
HOME
to confirm and save,
FORWARD
or
BACK
to scroll the pages.