7.7.
DISASSEMBLING THE GUARDS
Figure 7.7-1
Turn off the machine and unplug the mains lead
from the wall outlet.
1.
Back off the 5 screws
(a)
, using a 4mm
hexagonal wrench, on both sides of the
machine.
During reassembly, insert the longer
screws in the position indicated by the
arrows.
2.
Lift up the upper guard
(b)
securing it to the
upper side of the machine with some
adhesive tape or with a cord.
Figure 7.7-2
3.
Back off one of the 2 screw
(c)
, using a 5mm
hexagonal wrench.
4.
Remove the two front guards.
Figure 7.7-3
5.
Lower both the pedals.
6.
Back off the 3 screws
(d)
using a 5mm
hexagonal wrench.