background image

MiniBOSS O&M Manual Rev 3 

Page 7 

   9/25/2018 

 

2.7.

 

Pre-screening 

 

The  separator  is  not  designed  to  handle  heavy  solids.    A  Y-strainer  or  other  suitable  pre-
screening  filter  needs  to  be  installed  by  the  customer  in  front  of  the  separator  to  keep  larger 
particles  out  of  the  system

.   

A  5  micron  sediment  filter  is  included  in  the  system  to  handle 

some turbidity and to protect the oil/emulsion filters.    
 

3.

 

MAINTENANCE 

 

The system does not have installed redundancy.   The general maintenance philosophy for most 
of the components is to replace the components as they fail.  We have not developed a mean time 
between  failure  for  most  of  the  components  because  every  situation  is  different.    The  system 
does not generally run continuously—only when the bilge level increases to a designated level.  
In general, there is adequate time to make any required repair without impacting the bilge water 
level.  We recommend maintaining critical spare parts in inventory so that if there is a failure the 
part can be replaced quickly.  Every component can be changed within a few minutes.   
 
Components are fairly inexpensive and it is less expensive to replace the entire component than 
it  is  to  buy  the  part  and  overhaul the  component.    The  only  exception to  this  is  the  oil  content 
monitor,  which  is  the  most  expensive  component on  the  system.    In the  case  of  the  oil  content 
monitor,  it  is  against  IMO  regulations  for  the  customer  to  do  any  maintenance  on  the  system 
other than cleaning the cell.  If the monitor fails it generally has to replaced.  Any attempt by the 
customer to break the seal and open the monitor will void the validity of the monitor.   
 
There  are  some  routine  maintenance  functions  that  need  to  be  performed  to  keep  the  system 
running at peak performance:  

 

3.1.

 

Component Maintenance 

 

3.1.1.

 

Valve—The solenoid valve can be cleaned if it gets dirt in it. If the valve fails, it will need 
to be replaced with a new valve, it cannot be repaired..  
 

3.1.2.

 

OCM—The  OCM  will  need  a  periodic  check  of  zero  against  clean  water,  and  frequent 
cleaning of the glass sample cell.   The OCM measuring cell has a life of 5 years.  At the 
end of that period the measuring cell will need to be replaced. 

 

3.1.3.

 

Filters— The filters will need to be replaced when they become spent.   

 

3.1.4.

 

Controls components—The controls are minimal.  Any maintenance can be done when the 
component fails. 

 

3.1.5.

 

Pump—The pump is a diaphragm pump.   You cannot purchase individual components of 
the pump, and a repair kit is not available.  The pump is rated for continuous duty with a 
life expectancy of about 5,000 hours.  The pump will self-prime up to 9 feet vertical lift.     

   
 
 

Summary of Contents for MiniBOSS

Page 1: ...Wave MiniBOSS MEPC 107 49 110 240v AC 24v DC Monitored Oil Removal System OPERATION MAINTENANCE MANUAL Tel 44 1476 861717 info waveinternational co uk www waveinternational co uk...

Page 2: ...of the OCM 6 2 4 Changing the OCM Alarm Limit 6 2 5 Special tools 6 2 6 Shutdown procedures 6 2 7 Pre screening 7 3 MAINTENANCE 7 3 1 Component Maintenance 7 4 TROUBLE SHOOTING 8 4 1 Reduced flow thr...

Page 3: ...gm pump The design capacity of the standard MiniBOSS system is 1 gpm 0 25 M3 hr at 25 psig back pressure and no vacuum on the inlet or 1 5 gpm 0 35 M3 hr with the MiniBOSS 110 However the actual flow...

Page 4: ...s 83 lbs Capacity max 1 gpm 0 25 M3 hr 1 5 gpm 0 35 M3 hr Filter Media XP420 Extreme XP420 Extreme Test PSIG 55 55 Operating Pressure Up to 50 PSIG Up to 50 PSIG Operating Temp Range 1 55 C 1 55 C Oil...

Page 5: ...arm and Action Conditions The only alarm condition that requires operator action is when the oil removal filters have reached maximum hydrocarbon saturation This is determined by the inability of the...

Page 6: ...ither from your distributor or the factory to verify the calibration of the monitor It is not necessary to test the calibration as long as the cell has a valid calibration certificate The instructions...

Page 7: ...nent The only exception to this is the oil content monitor which is the most expensive component on the system In the case of the oil content monitor it is against IMO regulations for the customer to...

Page 8: ...of communication between the display and the detector loose cable c internal board or power supply failure See separate OCM O M manual for a more detailed explanation of the OCM 4 3 Solenoid Valve not...

Page 9: ...AKER 1P 2A C CURVE 41070 24V 110V GAUGE PRESSURE 100PSI BAR BRASS STEM GLYCOL 40371 24V 110V Measuring Cell Sensor for BilgMon 488 TPH 488MC 40456 24V 110V Cleaning Brush for BilgMon 488 Measuring Cel...

Reviews: