Chapter 3 –
TnT+
Portable Appliance
Part No: WCM-TNT+
Test Instructions
TnT+
Test Function Flow Chart
Menu 2
F1 Power Test
F2 Leakage Test
F3 RCD Test
Menu 2
F1 POWER TEST
Instructions for POWER TEST:
F1 + F3 = POWER RESULTS
Menu 2
RCD TEST
Instructions for RCD TEST for;
Time Test
O’ Phase - F3 + F1 + F1 + F3 = RESULTS
18O’ Phase - F3 + F1 + F2 + F3 = RESULTS
Change Current F3 + F1 + F3 + F3 = SETTING
Ramp Test F3 + F2 + F3 = RESULTS
Menu 2
F2 LEAKAGE TEST
Instructions for LEAKAGE TEST for;
•
Earthed Devices:
F2 + F1 + F3 = RESULTS
•
D.Ins., Ex-Lead:
F2 + F2 + F3 = RESULTS
•
RCD Protected:
F2 + F3 + F3 = RESULTS
APPLIANCE POWER TEST
Continue if Safe?
OK QUIT
Please note:
Menu 1 is common to all models [TnT, TnT-el & TnT+]
(refer page 14 for more details)
+
POWER RESULTS
240 V
0 W
0.000 A
0 VA
0.00 PF
LEAKAGE TEST
(F1) Earthed
Devices
(F2) D.Ins., Ex-Lead
To return to MENU 2
Press RETURN
EARTH LEAKAGE TEST
Continue if Safe?
OK QUIT
LEAKAGE RESULTS
PASS < 5.0 mA
0.0
mA
Pass ESC
EMERGENCY STOP
ABORT
To return to MENU 2
Press RETURN
LEAKAGE TEST
(F1) Earthed Devices
(F2) D.Ins., Ex-
Lead
EARTH LEAKAGE TEST
Continue if Safe?
OK QUIT
LEAKAGE RESULTS
PASS < 1.0 mA
0.2mA
Pass ESC
LEAKAGE TEST
(F1) Earthed Devices
(F2) D.Ins., Ex-Lead
(F3) RCD
Protected
EARTH LEAKAGE TEST
Continue if Safe?
OK QUIT
LEAKAGE RESULTS
PASS < 2.5 mA
1.0mA
Pass ESC
RCD TESTS
(F1) Time Test
(F2) Ramp Test
TIMED TEST 30mA
•
(F1) 0’ phase
•
(F2) 180’ phase
•
(F3) Change Current
RCD MULTIPLYER
•
(F1) Up 1.0
•
(F2) Down
•
(F3) START
RCD TIMED TEST
Time: 4ms
&
TIMED TEST 30mA
•
(F1) 0’ phase
•
(F2) 180’ phase
•
(F3) Change Current
RCD TEST CURRENT
•
(F1) Up 30mA
•
(F2) Down
•
(F3) SET
RCD TEST CURRENT
•
Up
30mA
•
Down
•
SAVING SETTINGS
RCD TESTS
(F1) Time Test
(F2) Ramp Test
RCD RAMP TEST
•
(F3) START
RCD RAMP TEST
TESTING
RCD RAMP TEST
Time: 4ms