3.2
MonitEM-Lab details
Below the unit, you can find:
• IEC connector for the main power cord.
• Ethernet connector.
• Output alarm connector (internal relay up to 240 Vac or 30 Vdc, 5A current,
normally closed).
• Sound button (pressed = sound on, unpressed = sound off).
• ON / OFF button (explained in more detailed in 4.3).
• Micro USB to connect to the PC.
• Reset switch.
• 1/4'' nut for tripod.
• 3 LEDs.
The 3 LEDs show the status of the device.
• The Ethernet LED is off when connection is established with the server.
• The status LED:
• Blinks twice per second when in measurement mode.
• Is lit continuously when connected to a computer.
• Blinks intermittently when in communication mode.
• Blinks rapidly for 5 seconds after initial calls (if communication is
successful).
• The LED of the SD indicates that it is writing. Do not switch off the device if
this LED is lit, since this could damage the SD.
3.3 Changing the probe
It is important to bear in mind that the memory of the unit is divided depending
on the probe type (
WPF3
,
WPF6
, etc...). Therefore, it is very important to
download all the data stored in the device before changing probes, in order to
avoid losing any information.
The measurement device must be switched off before the probe is changed.
Disconnect the power supply (mains cord) or use the switch off button. Once
User’s Manual
Electromagnetic Radiation Monitoring Device MonitEM-Lab
P.
6