57
A
dditional informa
tion
/
English
16.3
System disposal
Consult your local regulations when disposing of the
sensor insertion tool, the sensor guide, the transmitter,
and the charger.
Dispose of the sensor adhesive patch as medical waste.
16.4
Product specifications
16.4.1
Sensor specifications
Specification
Description
Glucose range
2.2–22.2 mmol/L
Sensor life
Up to 14 days
Calibration
GlucoMen Day METER (provided se-
parately) or commercially available
blood glucose meter.
Calibration range
3.9–13.9 mmol/L
Storage tempe-
rature
5°C to 35°C (41°F to 95°F)
Storage humidity
10% - 90% Relative humidity
Operating tempe-
rature
5°C to 45°C (41°F to 113°F)
Sterilization
Sterile by irradiation
Altitude
Up to 3048 metres (10,000 feet)
Atmospheric
pressure
700 hPa - 1060 hPa
16.4.2
Transmitter specifications
Specification
Description
Transmitter battery
type
Non-serviceable, rechargeable li-
thium-ion polymer battery
Transmitter battery
life
14 days of typical use
Transmitter life
5 years of average continuous use
Transmitter data
storage
14 days of calculated glucose data,
sampled once per minute
Transmitter dimen-
sions
Length: 33.0 mm, width: 24.0 mm,
thickness: 9.0 mm
Transmitter weight
5 grams
Ingress Protection
IP27: Protection against insertion
of large objects and immersion in 1
metre (3 feet) of water for up to 30
minutes.
Operating tempe-
rature
5°C to 45°C (41°F to 113°F)
Operating humidity
10% - 100% Relative humidity, con-
densing
Storage tempe-
rature
5°C to 45°C (41°F to 113°F)
Storage humidity
10% - 100% Relative humidity, con-
densing
Protection against
electrical shock
Type BF applied part
TX Frequency
2.402–2.480 GHz
Reading Frequency
Every 1 minute
Bandwidth
1.06 MHz
Maximum Output
Power
-2.93 dBm
Modulation
Gaussian FSK
Summary of Contents for A.MENARINI diagnostics GlucoMen Day CGM
Page 1: ...User Guide...
Page 65: ......