1.2.1 .12
Frequency Precision
1.2.2.3
Output
Dial Accuracy
(I
of s 1 % of full range) on x 100 thru
x 1 M ranges.
(2% of s
2%
of full range) on
x
.Ol
thru
x 10 and x 10M ranges.
Time Symmetry
0.5% on x 100 thru
x
IOOk ranges and from 0.2 to
2.0 on dial.
5% on all other ranges and from 0.02 to 2.0 on dial.
1.2.1 .13
Amplitude Precision
Amplitude Change With
Frequency
Sine variation less than:
0.1 dB thru x 1 OOk ranges;
0.5 dB on x 1 M range;
3 dB on
x
10M range.
Step Attenuator Accuracy
0.3 dB per 20 dB step at 2 kHz.
1.2.1.14
Waveform Characteristics
Sine Distortion
<0.5% on x 100 Hz to x 10 kHz.
< 1 .O% on x .01 to x 10 Hz and
x
100 kHz ranges.
All harmonics 30 dB below fundamental on
x
I MHz
range.
All harmonics 26 dB below fundamental on
x
10 MHz
range.
Square Wave Rise/Fall Times
At FUNCTION
OUT < 25
ns for 15V p-p into a
load.
Triangle Distortion
Odd harmonics within 15% of correct value to 2 MHz.
1.2.2
Modulation Generator
1.2.2.1
Waveforms
Selectable sine
, triangle
, square
, up
r a m p M a n d d o w n r a m p M .
1.2.2.2
Frequency Range
,
,
0.1 Hz to 100 kHz in three 100:1
ranges.
Sweep:
0.2 Hz
to
200
kHz (2
x
setting).
,
and
are fixed level 10V p-p balanced
about ground.
M
and
M
are fixed level 5 Vp
from 0 to + 5V.
1.2.2.4
Frequency Modulation (FM IN)
Voltage control of modulator frequency with sensitivity
of 20% of range per volt. Input impedance is 5
1.2.2.5
Waveform Characteristics
Sine Distortion
< 5 %
Time Symmetry
<
from 1
Hz
to 10 kHz.
from 0.1 Hz to 100 kHz.
1.2.3 General
1.2.3.1
Stability (for amplitude, dc offset and
frequency)
Short Term:
0.05% for 10 minutes.
Long Term:
for 24 hours.
1.2.3.2
Environmental
Specifications apply at 25°C
5 ° C ambient Instrument
will operate from O°C to 50°C ambient temperatures.
1.2.3.3
Dimensions
28.6
cm (11 in.) wide;
13.3
cm
(5
in.) high; 27.3 cm
in.) deep.
1.2.3.4
Weight
5 kg (11 lb) net; 6.6 kg
lb) shipping.
1.2.3.5
Power
90 to 105V, 108 to 126V, 198 to 231V and 216 to 252V
selectable; 48 to 400 Hz; less than 40 watts.
NOTE
Unless otherwise noted, ail specifications
apply from
0.1
to
2.0
on frequency dial
when FUNCTION OUT is at maximum and
terminated, with SYMMETRY control
at OFF. Symmetry and vernier affect fre-
quency calibration. Maximum possible
asymmetry is a function of frequency
setting.
1-3