4.3.20 TMx y...display/set trigger mode [0/1] trigger level [0/1]
Standard setting: 0 1
TM
provides parameter setting options for the auto-start trigger function which allows external triggering of the
auto-start command that was set via parameter AS. Triggering is accomplished via the external trigger input. All
starting modes which are selectable via AS can be launched and stopped by external triggering.
These are:
DS/DT/DW/DX/DF/DM/TP/LO/ID.
TM consists of two parameters which are separated by space (20 h).
X
0... Trigger function disabled
1... Trigger function enabled
y
0 ... low level of trigger signal starts measurement (HIGH
LOW transition)
1 ... high level of trigger signal starts measurement (LOW
HIGH transition)
For triggering, the trigger level must be permanently applied!
Examples:
a)
ASDT
TM1 1
Trigger signal = H
DT is performed
Trigger signal = L
DT is stopped
b)
ASDM
TM1 0
Trigger signal=H
no change in state
Trigger signal=L
DM active, i.e. one measurement is triggered
4.3.21 BRx...Display/set baudrate [2400...38400]
Standard setting: 9600
Available baud rate
BR
settings are: 2400, 4800, 9600, 19200, 38400. Faulty entries will be rounded to the
nearest baud rate. A fixed data format of eight data bits, with no partly and one stop bit is used.
Note:
After a change in baud rate setting, the communicating counterpart must also be set to the new baud rate.