5.4.

 

Tare Operation 

 
Tare is used if a tare weight is required to be removed from the gross weight, for example the weight of a 
container. 
 

Without weight applied ensure the display indicates “0” weight

 by pressing the 

[ON/ZERO]

 key. Place the 

container only upon the scale weighing pan, the weight of the container will be displayed. Press the 

[TARE]

 

key to remove this weight value and the display will indicate “0” but now the 

lamp over the 

TARE

 

wording will be lit. Now add the required load and the NET weight value will be displayed. 
 
When the load and tare weights are removed from the weighing pan the display will show a negative value 
and the lamps over the 

“ZERO” “TARE”

 wording will both be lit. Press the 

[TARE]

 key to remove the tare 

value. The scale will now display 

“0” and now only the 

lamp over the 

“ZERO” 

wording will be lit. 

 

5.5.

 

Save Power Function  

 
After 40 seconds of inactivity the scale will go to auto save power mode. To indicate the scale is in Auto 
save power mode the display will indicate a small 

“o”.

 

 
To wake the scale, add weight to the weighing pan. 
 

5.6.

 

Battery Charge Indication 

 
The battery charge level indicators are located to the left side of the display and show fully charge (higher 
than 6.3V), half charge (between 6V and 6.3V) and low charge (less than 6V).  
 

5.7.

 

Battery Charge 

 
Power to the scale is supplied by an internal rechargeable lead acid battery (6V/4.5Ah). When the display it 

indicates “

-bAtLo- 

“then

 the battery requires recharging. The 240V power lead for charging is located by 

sliding the cover plate from the base of the scale. Connect the 240V power lead to the supply and switch 
on, you may continue using the scale if required or switch it off for full recharging.  

 

5.8.

 

Set Parameter 

 
To display the different parameters of the scale Press and hold the 

[SET/0~9]

 key for 3 seconds to show the 

menu, then release the key.  
To step throught the parameter options press the 

[SET/0~9]

 key again to move to the next parameter. 

 
The parameter menu are as follows: 

 

rAngE (set limit)  

 

UnItS (choose unit)  

 

oFF (choose auto power off) 

 

FILt (set display mode)  

 

Zero (set auto zero)  

 

bUZZEr (set buzzer)  

 

d (choose division) 

 

on-rA (choose range of power on) 

 

d-dP (choose single or dual display) 

 
Press 

[SET/0~9] 

key is the select key

,  

Press 

[TARE]

 key is to confirm,  

Press

 [TARE]

 key again to exit

 

Summary of Contents for WS202

Page 1: ...Operator Manual for WS202 Digital Bench Scale ...

Page 2: ...nd Function 5 4 2 Display Messages 6 5 Operations 6 5 1 Preparation 6 5 2 Switching the Scale On and Off 6 5 3 Weighing 6 5 4 Tare Operation 7 5 5 Save Power Function 7 5 6 Battery Charge Indication 7 5 7 Battery Charge 7 5 8 Set Parameter 7 5 9 Units Change unit 9 5 10 To exit counting mode 9 5 11 A oFF Choose auto power off 9 5 12 FILtX Change display mode 9 5 13 Zero Set range of auto zero 10 5...

Page 3: ...e moving technique and ensure that The scale is switched off and disconnected from power before relocation A suitable trolley or other or other suitable means is used for the relocation Ensure that the new location is suitable for the task 1 3 General Operating Recommendations The following recommendations should be noted when operating the scale Do not overload the scale as this will damage the l...

Page 4: ...ontact your local council for recycling options 1 8 Correct Use Ensure that staff that will be using the WS202 scale are instructed in the correct use of the instrument and are made aware of the above safety instructions Incorrect use placement or handling of the scale may result in damage and possible injury 2 Inspection Instructions 2 1 Initial Inspection Upon receiving your new WS202 scale you ...

Page 5: ... C Internal battery 6V4 5Ah rechargeable battery 4 Keys and Display Messages 4 1 Keys and Function KEY NAME FUNCTION ON ZERO Press to turn the scale on Press to zero the scale Long press to enter LED power setting mode TARE Press to operate tare Press to confirm Press to exit Long press with power on to enter Calibration OFF Press to power off the scale SET Press to enter parameters Press to selec...

Page 6: ... the ON ZERO key the software version number will be displayed followed by the battery level voltage dc x xx The scale will then start a display check running through from 999999 to 000000 all the indicator lamps will light then go out Then the ZERO indication lamp will light and the display shown 0 Now the scale is read to use Switch off Press the OFF key to power off the instrument Auto off Afte...

Page 7: ...mall o To wake the scale add weight to the weighing pan 5 6 Battery Charge Indication The battery charge level indicators are located to the left side of the display and show fully charge higher than 6 3V half charge between 6V and 6 3V and low charge less than 6V 5 7 Battery Charge Power to the scale is supplied by an internal rechargeable lead acid battery 6V 4 5Ah When the display it indicates ...

Page 8: ...ext lowest move one unit right set this in the same way if required Press the TARE key until it has moved all the way through the units until the Low indicator light has gone out and now the HI indicator light is on Set the Hi upper level setting in the same way as the Low limit Step through the display digits until the Hi indicator light goes out and on is again shown on the display The scale wil...

Page 9: ...he sample number press SET 0 9 key to add 1 to the flashing unit number or move to next lowest digit by pressing the TARE key Press the SET 0 9 key moving down through the digits untill the scale returns to count mode The unit count is displated Remove sample load and add load to be counted The display will indicate the count Note The greater the number of the sample units the greater the accuracy...

Page 10: ... capacity 100 100 capacity Press the TARE key to confirm 5 17 d dP choose single or dual display When menu displays d dP press the TARE key to enter press the SET 0 9 key to choose no or yes no Operator display only single display yes both front and back displays working dual display Press the TARE key to confirm 5 18 Display LED Power Setting Mode With scale powered on Press the ON ZERO key until...

Page 11: ...100 maximum capacity 9e Remove some or all of weight bATLo is displayed The battery voltage is low Battery requires recharging Display reading not stable The scale is in motion Wait until the scale is stable The load is not stable Wait until the load is stable Large error in weight reading The scale is not zeroed before applying load Zero the scale before loading Requires re calibration See sectio...

Page 12: ... Operator Manual Page 12 of 12 WS202 Digital Bench Scale 8 Wedderburn Contact For any further enquiries please contact your local Wedderburn branch Phone 0800 800 379 Fax 0800 933 337 Website www wedderburn co nz ...