Repair information and repair instructions
25
7.6 Replacing the main circuit board
1. Open the unit (chapter 7.4, page 23).
Remove the main circuit board
1. Pull the plug
X200
and
X400
from the main
circuit board.
2. Remove the flat cables from front panel foil
and display:
To do so, pull the lock. Once you have pulled
the lock, you can remove the flat cable.
3. Remove pressure measurement tube
from
the pressure sensor or pull out the tube adapter
of the pressure measurement tube.
4. der Winkeltülle auf dem Drucksensor.
5. Remove the main circuit board
.
Inserting the main circuit board
1. Attach tube
to the pressure sensor of a new
main circuit board
adapter of the pressure measurement tube to
the angular bush on the pressure sensor.
2. Insert the main circuit board.
Flat cable
Lock
X200
X400
SOMNO
comfort
2/SOMNO
comfort
2e
Flat cable
Lock
X200
X400
SOMNO
balance
/SOMNO
balance
e
Flat cable
Lock
X200
X400
SOMNO
comfort
2/SOMNO
comfort
2e