Hygiene treatment
29
EN
6.2 Breathing tube
Caution!
The breathing tube may only be used again after cleaning if it is completely dry. If
moisture gets into the therapy device, this can result in damage to the device and
thus a hazard to users and patients.
Clean breathing tube
1. Pull the breathing tube off the device and off the exhalation system.
2. Seal both sides of the pressure measuring tube with a sealing plug.
3. Clean the creased tube and the adapter with a little detergent in hot water and make sure
no residues are left behind. Flush the inside of tube through thoroughly in the process.
4. Then rinse through thoroughly inside and out using clean hot water.
5. Shake the breathing tube thoroughly.
6. Hang up the breathing tube and allow it to drip-dry.
7. Dry the breathing tube with the aid of the therapy device.
Monthly
Replace fine filter (depending on contamination,
but in any event after no more than 250 hours or
one month (filter change indicator)).
Clean breathing tube.
Every 6 months
Replace coarse dust filter
Change pressure measuring tube
Every 12 months
Replace breathing mask and exhalation system.
Replace breathing tube.
Replace headgear/headband.
As required
Disinfect the SOMNO
vent
CR
Disinfect the SOMNO
click
300
See
associated
instructions
for use
When changing
patient
If necessary, hygiene treatment
Change particle filter of the bacteria filter.
See
associated
instructions
for use